Module qc tools
Project description
module-qc-tools v2.1.0
A general python tool for running ITkPixV1.1 module QC tests
Table of contents
- Requirements
- Installation
- Usage
- Configuration and external commands
- Measurements
- Output data
- Schema check
- Time Estimates
- Upload results to localDB
- For developer
Requirements
This tool requires users to have >= python3.7 with the following packages installed:
numpy
tabulate
Check the local python version with python -V
. If the local python version is
lower than Python 3.7, set up a virtual python environment following the
instructions here.
In addition, users shall install YARR
and prepare external scripts for remote
control of power supply, multimeter and NTC readout. The external scripts shall
follow a fixed format, which is detailed in the later section.
Installation
This package may be accessed by cloning from gitlab or by installing it via pip.
via clone
Use this method if you want to use the latest version of the package from gitlab. First clone the project:
git clone https://gitlab.cern.ch/atlas-itk/pixel/module/module-qc-tools.git
Upon a successful checkout, cd
to the new module-qc-tools
directory and run
the following to install the necessary software:
$ python -m venv venv
$ source venv/bin/activate
$ python -m pip install --upgrade pip
$ python -m pip install -e .
via pip
Use this method if you want to use the latest stable (versioned) release of the package.
python -m venv venv
source venv/bin/activate
python -m pip install -U pip
python -m pip install -U pip module-qc-tools==2.1.0
Usage
After installation, one just needs to enter the virtual environment in each new session to use the scripts:
source venv/bin/activate
The first step in using this package is to set up the Configuration and external commands input json file. This json file is then passed as input to the scripts used to run the Measurements.
Configuration and external commands
All the configuration/settings are defined in a single json file. The layer information needed for the different power settings is extracted from the module's serial number. Examples are provided in `$(module-qc-tools --prefix)/configs/':
example_merged_vmux.json
example_separate_vmux.json
$(module-qc-tools --prefix)/configs/example_merged_vmux.json
{
"yarr": {
"run_dir": "../Yarr",
"controller": "configs/controller/specCfg-rd53b-16x1.json",
"scanConsole_exe": "./bin/scanConsole",
"write_register_exe": "./bin/write-register",
"read_adc_exe": "./bin/read-adc",
"switchLPM_exe": "./bin/switchLPM",
"lpm_digitalscan": "configs/scans/rd53b/lpm_digitalscan.json",
"read_ringosc_exe": "./bin/rd53bReadRingosc"
},
"power_supply": {
"run_dir": "../labRemote",
"on_cmd": "./build/bin/powersupply -e ./src/configs/input-hw.json -n PS -c 1 power-on",
"off_cmd": "./build/bin/powersupply -e ./src/configs/input-hw.json -n PS -c 1 power-off",
"set_cmd": "./build/bin/powersupply -e ./src/configs/input-hw.json -n PS -c 1 set-current {i} {v}",
"getI_cmd": "./build/bin/powersupply -e ./src/configs/input-hw.json -n PS -c 1 meas-current",
"getV_cmd": "./build/bin/powersupply -e ./src/configs/input-hw.json -n PS -c 1 meas-voltage",
"n_try": 0,
"success_code": 0
},
"multimeter": {
"run_dir": "../labRemote",
"dcv_cmd": [
"./build/bin/meter -e ./src/configs/input-hw-min_marija.json -n myVmuxMeter -c 0 meas-voltage"
],
"n_try": 0,
"success_code": 0
},
"ntc": {
"run_dir": "../labremote_devel/scripts",
"cmd": "python measureT.py",
"n_try": 0,
"success_code": 0
},
"tasks": {
"GENERAL": {...},
"ADC_CALIBRATION": {...},
"ANALOG_READBACK": {...},
"SLDO": {...},
"VCAL_CALIBRATION": {...},
"INJECTION_CAPACITANCE": {...}
}
}
The major blocks (e.g. yarr
, power_supply
, multimeter
, ntc
) correspond
to how the scripts will communicate with the module via YARR and how they will
communicate with the lab equipment. Each of these blocks are explained in the
following sections. The task
block specifies the settings of each scan
performed by the scripts, and will be explained in the
Measurements section.
yarr
The yarr
block specifies the path to the YARR
repository as well as the
corresponding YARR configuration files.
run_dir
: path (relative or absolute) to the directory whereYARR
commands should be runcontroller
: path (relative torun_dir
or absolute) to the controller filescanConsole_exe
: path (relative torun_dir
or absolute) to thescanConsole
executablewrite_register_exe
: path (relative torun_dir
or absolute) to thewrite_register
executableread_register_exe
: path (relative torun_dir
or absolute) to theread_register
executableread_adc_exe
: path (relative torun_dir
or absolute) to theread_adc
executableswitchLPM_exe
: path (relative torun_dir
or absolute) to theswitchLPM
executablelpm_digitalscan
: path (relative torun_dir
or absolute) to the low-power mode digital scan in YARRread_ringosc_exe
: path (relative torun_dir
or absolute) to theread_ringosc_exe
executablesuccess_code
: exit status that indicates success. The default is 0 besidesconfigure
, which has exit status 1 for success. The user do not need to set thesuccess_code
as the default in QC software is synchronized with YARR.
power_supply
The power_supply
block specifies the path and the commands for handling the
low voltage power supply
run_dir
: path (relative or absolute) to the directory wherepower_supply
commands should be runon_cmd
: command to turn on the power supply with specified voltage and current. Use the syntax{v}
and{i}
to represent the voltage and current that are to be given as input argumentsoff_cmd
: command to turn off the power supplyset_cmd
: command to set voltage and current for power supply. Use the syntax{v}
and{i}
to represent the voltage and current that are to be given as input argumentsgetI_cmd
: command to measure current of the power supply. This command shall return a std output which represents the value of the current (float in the unit of [A]). For example, when I = 5.2A,getI_cmd
returns std output =5.2
.getV_cmd
: command to measure voltage of the power supply. This command shall return a std output which represents the value of the voltage (float in the unit of [V]). For example, when V = 1.8V,getV_cmd
returns std output =1.8
.n_try
: number of re-tries in case the script fails to read from the power supplysuccess_code
: exit status that indicates success. The default is 0.
high_voltage
Similar to the power_supply
block, the high_voltage
block specifies the path
and the commands for handling the high voltage power supply needed for leakage
current vs. bias voltage (IV) measurements.
ramp_cmd
: (optional) command to ramp the high voltage to the target value without measuring. If no command is available or provided, module-qc-tools will use its internal ramp function.getI_cmd
: command to measure the leakage current of the high voltage. This command shall return a std output which represents the value of the current (float in the unit of [A]). For example, when I = 1uA,getI_cmd
returns std output =0.000001
.
multimeter
The multimeter
block specifies the path and the commands for handling the
multimeter
run_dir
: path (relative or absolute) to the directory wheremultimeter
commands should be rundcv_cmd
: list of commands to measure voltage from the multimeter. Each command corresponds to a single multimeter channel (only the used channels need to be listed). Each command returns a std output which represents the value of measured voltage (float in the unit of [V]). For example, when V = 0.352V,dcv_cmd
returns std output =0.352
.n_try
: number of re-tries in case the script fails to read from the multimetersuccess_code
: exit status that indicates success. The default is 0.
ntc
The ntc
block specifies the path and the commands for handling the NTC
run_dir
: path (relative or absolute) to the directory wherentc
commands should be runcmd
: command to measure temperature from the module NTC. The command returns a std output which represents the value of measured temperature (float in the unit of [C]). For example: when T = 36.2C,cmd
returns std output =36.2
.n_try
: number of re-tries in case the script fails to read from the ntcsuccess_code
: exit status that indicates success. The default is 0.
tasks
The tasks
block starts with the GENERAL
section that specifies the
layer-dependent power settings:
v_max
: the voltage to be set to the power supply (i.e. the max voltage since the power supply should operate in constant current)i_config
: the current at which the module should be configuredshare_vmux
: whether to merge the Vmux channels or notv_mux_channels
: multimeter channel to measure the Vmux for each chip (correspond to each element in the dcv_cmd in the multimeter block)
The main part of the tasks
block is to specify all
measurements as listed below.
Measurements
An overview of the steps in the module QC procedure is documented in the Electrical specification and QC procedures for ITkPixV1.1 modules document and in this spreadsheet. Each measurement is performed with one script. All scripts assume that the modules to be tested are already powered on.
IV MEASURE
measurement-IV-MEASURE
This script will test the sensor leakage current vs. reverse bias voltage
(task = IV_MEASURE
) as specified in the input configuration json file (i.e.
$(module-qc-tools --prefix)/configs/example_merged_vmux.json
).
Configuration settings
v_min
: the starting voltage of this measurementv_max
: the end voltage of this measurementi_max
: the current compliance throughout the measurementn_points
: how many points should be measured depending on the required voltage steps (i.e. 1V for 3D and 5V for planar modules)settling_time
: delay in seconds between setting a bias voltage and reading the current
Help message
$ measurement-IV-MEASURE --help
usage: measurement-IV-MEASURE [-h] [-c CONFIG] [-o OUTPUT_DIR] [-m MODULE_CONNECTIVITY] [--verbosity LEVEL] [--site STRING] [-t FLOAT] [--vdepl FLOAT]
optional arguments:
-h, --help show this help message and exit
-c CONFIG, --config CONFIG
Config file
-o OUTPUT_DIR, --output-dir OUTPUT_DIR
output directory
-m MODULE_CONNECTIVITY, --module-connectivity MODULE_CONNECTIVITY
path to the module connectivity. Used also to identify the module SN, and to set the default output directory.
-v VERBOSITY, --verbosity VERBOSITY
Log level [options: DEBUG, INFO (default) WARNING, ERROR]
--site STR Your testing site. Required when submitting results to the database. Please use institute codes defined on production DB, i.e. "LBNL_PIXEL_MODULES" for LBNL, "IRFU" for Paris-Saclay, ...
--vdepl FLOAT Depletion voltage from production database [default: None]
Example
measurement-IV-MEASURE -c $(module-qc-tools --prefix)/configs/example_merged_vmux.json -m ~/module_data/20UPGR91301046/20UPGR91301046_L2_LP.json
Example command to run on the toy emulator
measurement-IV-MEASURE -c $(module-qc-tools --prefix)/configs/emulator_merged_vmux.json -o emulator/outputs
ADC calibration
measurement-ADC-CALIBRATION
This script will run the ADC calibration (task = ADC_CALIBRATION
) as specified
in the input configuration json file (i.e.
$(module-qc-tools --prefix)/configs/example_merged_vmux.json
).
Configuration settings
MonitorV
: list of Vmux channels to be measuredInjVcalRange
: the range of the calibration injection circuit(1: a large range and 0: a small range i.e. half the large range but a finer step)Range
: the DACs scan range ["start", "stop", "step"]
Help message
$ measurement-ADC-CALIBRATION --help
usage: measurement-ADC-CALIBRATION [-h] [-c CONFIG] [-o OUTPUT_DIR] [--verbosity LEVEL] [-m MODULE_CONNECTIVITY] [--perchip]
optional arguments:
-h, --help show this help message and exit
-c CONFIG, --config CONFIG
Config file
-o OUTPUT_DIR, --output-dir OUTPUT_DIR
output directory
-m MODULE_CONNECTIVITY, --module-connectivity MODULE_CONNECTIVITY
path to the module connectivity. Used also to identify the module SN, and to set the default output directory
-v VERBOSITY, --verbosity VERBOSITY
Log level [options: DEBUG, INFO (default) WARNING, ERROR]
--perchip Store results in one file per chip (default: one file per module)
--site, Your testing site. Required when submitting results to the database. Please use institute codes defined on production DB, i.e. "LBNL_PIXEL_MODULES" for LBNL, "IRFU" for Paris-Saclay, ...
Example
measurement-ADC-CALIBRATION -c $(module-qc-tools --prefix)/configs/example_merged_vmux.json -m ~/module_data/20UPGR91301046/20UPGR91301046_L2_warm.json
Example command to run on the toy emulator
measurement-ADC-CALIBRATION -c $(module-qc-tools --prefix)/configs/emulator_merged_vmux.json -o emulator/outputs/
Analog readback
measurement-ANALOG-READBACK
This script will measure all internal voltages available through VMUX and IMUX,
measure the chip temperature, and measure VDDA/VDDD/ROSC vs Trim. channels. The
scan settings are defined in the task = ANALOG_READBACK
block of the input
configuration file (i.e.
$(module-qc-tools --prefix)/configs/example_merged_vmux.json
). The NTC needs
to be set in order to run this script, so that the temperature can be read.
Configuration settings
v_mux
: list of Vmux channels to be measuredi_mux
: list of Imux channels to be measuredv_mux_ntc
: list of Vmux channels to be measured for ntc temperaturei_mux_ntc
: list of Imux channels to be measured for ntc temperaturev_mux_tempsens
: list of Vmux channels to be measured for 3 temperature sensorsMonSensSldoDigSelBias
: Bias 0 and 1 for MOS sensor near digital SLDOMonSensSldoAnaSelBias
: Bias 0 and 1 for MOS sensor near analog SLDOMonSensAcbSelBias
: Bias 0 and 1 for MOS sensor near centerMonSensSldoDigDem
: Dem 0-15 for MOS sensor near digital SLDOMonSensSldoAnaDem
: Dem 0-15 for MOS sensor near analog SLDOMonSensAcbDem
: Dem 0-15 for MOS sensor near centerSldoTrimA
: Sldo analog Trim 0-15SldoTrimD
: Sldo digital Trim 0-15
Help message
usage: measurement-ANALOG-READBACK [-h] [-c CONFIG] [-o OUTPUT_DIR]
[-m MODULE_CONNECTIVITY] [--verbosity LEVEL] [--perchip]
optional arguments:
-h, --help show this help message and exit
-c CONFIG, --config CONFIG
Config file
-o OUTPUT_DIR, --output-dir OUTPUT_DIR
output directory
-m MODULE_CONNECTIVITY, --module-connectivity MODULE_CONNECTIVITY
path to the module connectivity. Used also
to identify the module SN, and to set the default output directory
-v VERBOSITY, --verbosity VERBOSITY
Log level [options: DEBUG, INFO (default) WARNING, ERROR]
--perchip Store results in one file per chip (default: one file per module)
--site, Your testing site. Required when submitting results to the database. Please use institute codes defined on production DB, i.e. "LBNL_PIXEL_MODULES" for LBNL, "IRFU" for Paris-Saclay, ...
Example
measurement-ANALOG-READBACK -c $(module-qc-tools
--prefix)/configs/example_merged_vmux.json -m
~/module_data/20UPGR91301046/20UPGR91301046_L2_warm.json
Example command to run on the toy emulator
measurement-ANALOG-READBACK -c $(module-qc-tools
--prefix)/configs/emulator_merged_vmux.json -o emulator/outputs
SLDOVI
measurement-SLDO
This script will run the VI scans (task = SLDO
) as specified in the input
configuration json file (i.e.
$(module-qc-tools --prefix)/configs/example_merged_vmux.json
).
Configuration settings
i_min
: the minimum current of the VI scani_max
: the maximum current of the VI scann_points
: how many points should be measured (equally spread between the max of the min currents)extra_i
: extra current points to be measuredv_mux
: list of Vmux channels to be measuredi_mux
: list of Imux channels to be measured
Help message
$ measurement-SLDO --help
usage: measurement-SLDO [-h] [-c CONFIG] [-o OUTPUT_DIR] [-m MODULE_CONNECTIVITY] [--perchip]
[--verbosity LEVEL]
optional arguments:
-h, --help show this help message and exit
-c CONFIG, --config CONFIG
Config file
-o OUTPUT_DIR, --output-dir OUTPUT_DIR
output directory
-m MODULE_CONNECTIVITY, --module-connectivity MODULE_CONNECTIVITY
path to the module connectivity. Used also to identify the module SN, and to
set the default output directory
--perchip Store results in one file per chip (default: one file per module)
-v VERBOSITY, --verbosity VERBOSITY
Log level [options: DEBUG, INFO (default) WARNING, ERROR]
--site, Your testing site. Required when submitting results to the database. Please use institute codes defined on production DB, i.e. "LBNL_PIXEL_MODULES" for LBNL, "IRFU" for Paris-Saclay, ...
Example
measurement-SLDO -c $(module-qc-tools --prefix)/configs/example_merged_vmux.json -m ~/module_data/20UPGR91301046/20UPGR91301046_L2_warm.json
Example command to run on the toy emulator
measurement-SLDO -c $(module-qc-tools --prefix)/configs/emulator_merged_vmux.json -o emulator/outputs
VCal calibration
measurement-VCAL-CALIBRATION
This script will run the VCal calibration (task = VCAL_CALIBRATION
) as
specified in the input configuration json file (i.e.
$(module-qc-tools --prefix)/configs/example_merged_vmux.json
).
Configuration settings
InjVcalRange
: the range of the calibration injection circuit(1: a large range and 0: a small range i.e. half the large range but a finer step)MonitorV
: two DACs VMUX assignments Vcal_med(8) and Vcal_high(7)MonitorV_GND
: the GNDA VMUX assignment 30Large_Range
: the large DACs scan range ["start", "stop", "step"]Small_Range
: the small DACs scan range ["start", "stop", "step"]
Help message
$ measurement-VCAL-CALIBRATION --help
usage: measurement-VCAL-CALIBRATION [-h] [-c CONFIG] [-o OUTPUT_DIR] [-m MODULE_CONNECTIVITY] [--verbosity LEVEL] [--perchip]
optional arguments:
-h, --help show this help message and exit
-c CONFIG, --config CONFIG
Config file
-o OUTPUT_DIR, --output-dir OUTPUT_DIR
output directory
-m MODULE_CONNECTIVITY, --module-connectivity MODULE_CONNECTIVITY
path to the module connectivity. Used also to identify the module SN, and to
set the default output directory
-v VERBOSITY, --verbosity VERBOSITY
Log level [options: DEBUG, INFO (default) WARNING, ERROR]
--perchip Store results in one file per chip (default: one file per module)
--site, Your testing site. Required when submitting results to the database. Please use institute codes defined on production DB, i.e. "LBNL_PIXEL_MODULES" for LBNL, "IRFU" for Paris-Saclay, ...
Example
measurement-VCAL-CALIBRATION -c $(module-qc-tools
--prefix)/configs/example_merged_vmux.json -m
~/module_data/20UPGR91301046/20UPGR91301046_L2_warm.json
Example command to run on the toy emulator
measurement-VCAL-CALIBRATION -c $(module-qc-tools
--prefix)/configs/emulator_merged_vmux.json -o emulator/outputs/
Injection capacitance
measurement-INJECTION-CAPACITANCE
This script will run the injection capacitance measurement
(task = INJECTION_CAPACITANCE
) as specified in the input configuration json
file (i.e. $(module-qc-tools --prefix)/configs/example_merged_vmux.json
).
Configuration settings
n_meas
: number of measurements to performv_mux
: list of Vmux channels to be measuredi_mux
: list of Imux channels to be measured
Help message
usage: measurement-INJECTION-CAPACITANCE [-h] [-c CONFIG]
[-o OUTPUT_DIR] [-m MODULE_CONNECTIVITY] [--verbosity]
[--perchip]
optional arguments:
-h, --help show this help message and exit
-c CONFIG, --config CONFIG
Config file
-o OUTPUT_DIR, --output-dir OUTPUT_DIR
output directory
-m MODULE_CONNECTIVITY, --module-connectivity MODULE_CONNECTIVITY
path to the module connectivity. Used also
to identify the module SN, and to set the default output directory
-v VERBOSITY, --verbosity VERBOSITY
Log level [options: DEBUG, INFO (default) WARNING, ERROR]
--perchip Store results in one file per chip (default: one file per module)
--site, Your testing site. Required when submitting results to the database. Please use institute codes defined on production DB, i.e. "LBNL_PIXEL_MODULES" for LBNL, "IRFU" for Paris-Saclay, ...
Example
measurement-INJECTION-CAPACITANCE -c $(module-qc-tools
--prefix)/configs/example_merged_vmux.json -m
~/module_data/20UPGR91301046/20UPGR91301046_L2_warm.json
Example command to run on the toy emulator
measurement-INJECTION-CAPACITANCE -c $(module-qc-tools
--prefix)/configs/emulator_merged_vmux.json -o emulator/outputs/
Low Power Mode
measurement-LP-MODE
This script will run the low power mode test (task = LP_MODE
) as specified in
the input configuration json file (i.e.
$(module-qc-tools --prefix)/configs/example_merged_vmux.json
).
Configuration settings
v_max
: the voltage to be set to the power supply specific to this measurementi_config
: the current at which the module should be configured specific for this measurementv_mux
: list of Vmux channels to be measuredi_mux
: list of Imux channels to be measured
Help message
usage: measurement-LP-MODE [-h] [-c CONFIG] [-o OUTPUT_DIR]
[-m MODULE_CONNECTIVITY] [--verbosity LEVEL] [--perchip]
optional arguments:
-h, --help show this help message and exit
-c CONFIG, --config CONFIG
Config file
-o OUTPUT_DIR, --output-dir OUTPUT_DIR
output directory
-m MODULE_CONNECTIVITY, --module-connectivity MODULE_CONNECTIVITY
path to the module connectivity. Used also
to identify the module SN, and to set the default output directory.
Note that this measurement should be used using LP config therefore
this should point to LP connectivity.
-v VERBOSITY, --verbosity VERBOSITY
Log level [options: DEBUG, INFO (default) WARNING, ERROR]
--perchip Store results in one file per chip (default: one file per module)
--site, Your testing site. Required when submitting results to the database. Please use institute codes defined on production DB, i.e. "LBNL_PIXEL_MODULES" for LBNL, "IRFU" for Paris-Saclay, ...
Example
measurement-LP-MODE -c $(module-qc-tools --prefix)/configs/example_merged_vmux.json -m ~/module_data/20UPGR91301046/20UPGR91301046_L2_LP.json
Example command to run on the toy emulator
measurement-LP-MODE -c $(module-qc-tools --prefix)/configs/emulator_merged_vmux.json -o emulator/outputs/
Overvoltage Protection
measurement-OVERVOLTAGE-PROTECTION
This script will test the Overvolatge Protection (OVP)
(task = OVERVOLTAGE_PROTECTION
) as specified in the input configuration json
file (i.e. $(module-qc-tools --prefix)/configs/example_merged_vmux.json
).
Configuration settings
v_max
: the voltage to be set to the power supply specific to this measurementi_config
: the current at which the module should be configured specific for this measurementv_mux
: list of Vmux channels to be measuredi_mux
: list of Imux channels to be measured
Help message
$ measurement-OVERVOLTAGE-PROTECTION --help
usage: measurement-OVERVOLTAGE-PROTECTION [-h] [-c CONFIG] [-o OUTPUT_DIR] [-m MODULE_CONNECTIVITY] [--perchip] [--verbosity LEVEL]
optional arguments:
-h, --help show this help message and exit
-c CONFIG, --config CONFIG
Config file
-o OUTPUT_DIR, --output-dir OUTPUT_DIR
output directory
-m MODULE_CONNECTIVITY, --module-connectivity MODULE_CONNECTIVITY
path to the module connectivity. Used also to identify the module SN, and to set the default output directory.
Note that this measurement should be used using LP config therefore
this should point to LP connectivity.
--perchip Store results in one file per chip (default: one file per module)
-v VERBOSITY, --verbosity VERBOSITY
Log level [options: DEBUG, INFO (default) WARNING, ERROR]
--site, Your testing site. Required when submitting results to the database. Please use institute codes defined on production DB, i.e. "LBNL_PIXEL_MODULES" for LBNL, "IRFU" for Paris-Saclay, ...
Example
measurement-OVERVOLTAGE-PROTECTION -c $(module-qc-tools --prefix)/configs/example_merged_vmux.json -m ~/module_data/20UPGR91301046/20UPGR91301046_L2_LP.json
Example command to run on the toy emulator
measurement-OVERVOLTAGE-PROTECTION -c $(module-qc-tools --prefix)/configs/emulator_merged_vmux.json -o emulator/outputs
Output data
The output of the measurements follows the structure below:
|Measurements
|<test_type>
|VCAL_CALIB
|<timestamp>
|<YYYY-MM-DD_HHMMSS>
|<chip_name/module_serial_number>.json
The test-type of each measuremnt is the corresponding test-code used in the Production Data Base. The naming of each measuremnet script is chosen to be the same as the test-type. The timestamp is chosen to be the start time of the measurement.
Schema check
The schema for the output json files is checked based on the schema files
specified in the folder schema
. To run the common schema check for all test
outputs, do:
jsonschema --instance [path to output json files] $(module-qc-tools
--prefix)/schema/common.json
To run the further schema check for a specific test output, do:
jsonschema --instance [path to output json files] $(module-qc-tools
--prefix)/schema/[qc_task].json
Example (with the emulator output files)
jsonschema --instance emulator/outputs/SLDO_reference/<timestamp>/chip1.json
$(module-qc-tools --prefix)/schema/common.json jsonschema --instance
emulator/outputs/SLDO_reference/<timestamp>/chip1.json $(module-qc-tools
--prefix)/schema/SLDO.json
Time Estimates
Measurement | Duration |
---|---|
ADC calib | 00:02:12 |
Analog Readback | 00:18:27 |
SLDO VI | 00:08:45 |
Vcal calib | 00:04:50 |
Injection Cap. | 00:00:50 |
LP Mode | 00:00:53 |
OVP | 00:00:27 |
Upload results to localDB
The output measurement files can be uploaded to localDB using the following command:
module-qc-tools-upload --path PATH --host HOST --port PORT
Given a path to a directory with the output files, the script will recursively
search the directory and upload all files with the .json
extension. Supply the
option --dry-run
to see which files the script finds without uploading to
localDB.
Help message
$ module-qc-tools-upload -h
usage: module-qc-tools-upload [-h] [--path PATH] [--host HOST] [--port PORT] [-n]
Walk through the specified directory (recursively) and attempt to submit all json files to LocalDB as the QC measurement
optional arguments:
-h, --help show this help message and exit
--path PATH Path to directory with output measurement files
--host HOST localDB server, default is localhost
--port PORT localDB port, default is 5000
-n, --dry-run Dry-run, do not submit to localDB.
For Developer
versioning
In case you need to tag the version of the code, you need to have either hatch
or pipx
installed.
- Activate python environment, e.g.
source venv/bin/activate
. - Run
python -m pip install hatch
orpython -m pip install pipx
.
You can bump the version via:
pipx run hatch run tag x.y.z
# or
hatch run tag x.y.z
where x.y.z
is the new version to use. This should be run from the default
branch (main
/ master
) as this will create a commit and tag, and push for
you. So make sure you have the ability to push directly to the default branch.
pre-commit
Install pre-commit to avoid CI failure. Once pre-commit is installed, a git hook script will be run to identify simple issues before submission to code review.
Instruction for installing pre-commit in a python environment:
- Activate python environment, e.g.
source venv/bin/activate
. - Run
python3 -m pip install pre-commit
. - Run
pre-commit install
to install the hooks in.pre-commit-config.yaml
.
After installing pre-commit, .pre-commit-config.yaml
will be run every time
git commit
is done. Redo git add
and git commit
, if the pre-commit script
changes any files.
Project details
Release history Release notifications | RSS feed
Download files
Download the file for your platform. If you're not sure which to choose, learn more about installing packages.
Source Distribution
Built Distribution
File details
Details for the file module_qc_tools-2.1.0.tar.gz
.
File metadata
- Download URL: module_qc_tools-2.1.0.tar.gz
- Upload date:
- Size: 94.1 kB
- Tags: Source
- Uploaded using Trusted Publishing? No
- Uploaded via: twine/4.0.2 CPython/3.7.17
File hashes
Algorithm | Hash digest | |
---|---|---|
SHA256 | 83f23a24fd34c2e5be15fe90cac59166f5a0f71158796c137c4c4bab229351fe |
|
MD5 | add27b0d32b03f7fde90dc3ef91a62fa |
|
BLAKE2b-256 | 07eff113d2610102b57c171d400d2f93a68f9d3c15fc179cb2749329863b1c89 |
File details
Details for the file module_qc_tools-2.1.0-py3-none-any.whl
.
File metadata
- Download URL: module_qc_tools-2.1.0-py3-none-any.whl
- Upload date:
- Size: 100.6 kB
- Tags: Python 3