Skip to main content

Remote control of Rigol DS1000E/D oscilloscopes

Project description

Remote control of Rigol DS1000E/D series digital oscilloscopes.

Example usage:

>>> import matplotlib.pyplot as plt
>>> from ds1052 import DS1052, PointsMode
>>> with DS1052() as dso:
...     # Acquire the waveform of channel 1 in "normal" points mode
...     # which returns 600 data points.
...     waveform = dso.read_waveforms([1], PointsMode.normal)[0]
...
>>> # Vertical scale and offset.
>>> waveform.v_scale
1.0
>>> waveform.v_offset
0.52
>>> # Timebase scale and offset.
>>> waveform.tb_scale
0.0005
>>> waveform.tb_offset
0.0
>>> # waveform.t is a numpy array with time axis values. 0 is the trigger
>>> # time; waveform.v is a numpy array with the voltage values.
>>> plt.plot(waveform.t, waveform.v)
[<matplotlib.lines.Line2D object at 0x7f8b8e075640>]
>>> plt.show()
>>>

Most settings of the DS1000 oscilloscopes are available as properties:

>>> import ds1052
>>> dso = ds1052.DS1052()
>>> dso.open()
>>> dso.timebase_scale
0.0005
>>> dso.timebase_offset
0.0
>>> dso.trigger.edge.level
1.48
>>> dso.trigger.edge.coupling
<TriggerCoupling.dc: 'DC'>
>>> dso.channel[2].coupling
<ChannelCoupling.dc: 'DC'>
>>> dso.channel[2].scale
1.0
>>> dso.channel[2].offset
-3.52
>>> dso.channel[2].scale = 0.6
>>> dso.channel[2].scale
0.6

>>> # Measurement values are returned as tuples (value, qualifier).
>>> dso.channel[1].voltage_rms
(2.14, <MeasurementQualifier.value: ('value',)>)
>>> # MeasurementQualifier.less_than indicates that the DSO could make
>>> # a measurement with sufficient precision. (Settings of this example:
>>> # Channel 1 input is a 1kHz square wave; the sampling rate is 500k/s.
>>> # 500 samples per period are clearly not enough to measure the rise
>>> # time.)
>>> dso.channel[1].rise_time
(3e-05, <MeasurementQualifier.less_than: 'less than'>)

>>> # Finally, DS1052.close() should be called to unlock the keyboard.
>>> # (If DS1052 is used as a context manager, the close() call is made
>>> # when the context is left.)
>>> dso.close()

Aims of ya_ds1052

  1. Provide an easy to use remote control interface to DS1000E/D oscilloscopes.

  2. Provide, where possible, workarounds for the many quirks of the DS1000E/D, which have a, well, unconventional SCPI implementation. The most glaring problem is that the DSOs do not implement the command *OPC? (operation complete command). This makes it sometimes difficult, but at least cumbersome, to figure out when the change of a setting is eventually applied. This library tries, where possible, to implement checks if such changes have been fully applied and when an operation can be performed without slowing down the performance too drastically. (Rigol's own Ultrascope software sends SCPI commands with a minimum time distance of 0.12 seconds. This makes some operations like a read-out of all settings really slow and does not seems to be necessary.)

Installation:

pip install ya_ds1052

This installs the additional libraries numpy and aenum.

One important dependency must be installed manually in many cases: An implementation of the USBTMC layer. ya_ds1052 works with three USBTMC drivers:

  • The usbtmc driver of the Linux kernel. Usage of this driver requires no additional Python module.

  • The Python package python-usbtmc. This package in turn needs the PyUSB package. See http://alexforencich.com/wiki/en/python-usbtmc/start for more details. Note that this package has not been updated since a longer time.

  • The Python package PyVISA, together with a suitable backend. ya_ds1052 is tested with the backends pyvisa-py (another Python package) and with the National Instruments VISA library for Linux (a bit tricky to install for recent Linux kernel versions).

    Be aware that the USBTMC implementation of the DS1000E/D is not fully standard compliant. The VISA backend must provide some workarounds for the DS1000's quirks.

    Run pyvisa-info to check which backend is detected by PyVISA.

On Linux it can be necessary to add a udev rule to give users access to the device files for the DS1000 oscilloscope. Add a file named 45-ds1000.rules to the directory `/etc/udev/rules.d/' with the following content:

SUBSYSTEMS=="usb", ACTION=="add", ATTRS{idVendor}=="1ab1", ATTRS{idProduct}=="0588", GROUP="users", MODE="0660"

(Change the group name users to whatever suits your needs to manage access rights for different users.)

Then run

udevadm control --reload-rules && udevadm trigger

to activate the new rule. Ensure that you and other users who will use the oscilloscope are members of the user group specified in the udev rule. If you must add yourself or other users to the user group, remember that a new group memebership is recognized after a new login, i.e., users must logout and login again before access to the device files is possible.

Changelog

0.1.2

  • New method DS1052.screenshot(): Retrieves a BMP image from the DS1052.
  • Property DS1052.channel[].filter_enabled: Raise DS1052PropertySetError on attempts to set this property while the DS1052 is in "stop" mode.
  • Improved range check of DS1052.trigger.{edge,pulse,video}.level: Only allow values in the range (-6 * v_scale .. 6 * v_scale) - v_offset where v_scale and v_offset are the vertical scale/offset ot the tirgger source.
  • Same check as above for DS1052.trigger.slope.voltage_level_lower and DS1052.trigger.slope.voltage_level_upper.
  • classes TriggerModeSettingsEdge, TriggerModeSettingsPulse, TriggerModeSettingsVideo, TriggerModeSettingsSlope, TriggerModeSettingsAlternation, Trigger, Channel, _Channels, DS1052: Methods get_config() and set_config() added.
  • New properties Channel.offset_range, Channel.trigger_level_range Channel.scale_range
  • Property Channel.offset: Incorrect calculation of value limits fixed.

0.1.1

First public release

Project details


Download files

Download the file for your platform. If you're not sure which to choose, learn more about installing packages.

Source Distribution

ya_ds1052-0.1.2.tar.gz (90.2 kB view details)

Uploaded Source

File details

Details for the file ya_ds1052-0.1.2.tar.gz.

File metadata

  • Download URL: ya_ds1052-0.1.2.tar.gz
  • Upload date:
  • Size: 90.2 kB
  • Tags: Source
  • Uploaded using Trusted Publishing? No
  • Uploaded via: twine/3.1.1 pkginfo/1.4.2 requests/2.22.0 setuptools/45.2.0 requests-toolbelt/0.8.0 tqdm/4.30.0 CPython/3.8.10

File hashes

Hashes for ya_ds1052-0.1.2.tar.gz
Algorithm Hash digest
SHA256 b832334513dd40f10246cace07a2671d060ba7370909e0b678664abde1619ee5
MD5 833fbd5f63a827b7987b5b5844b4fd02
BLAKE2b-256 820e82b5bd6fc43b23b272330e3b8d2c154b14b0013db268f6a9c03b3c534d89

See more details on using hashes here.

Supported by

AWS AWS Cloud computing and Security Sponsor Datadog Datadog Monitoring Fastly Fastly CDN Google Google Download Analytics Microsoft Microsoft PSF Sponsor Pingdom Pingdom Monitoring Sentry Sentry Error logging StatusPage StatusPage Status page